This precision, patented UV sensor measures the sunburn-causing portion of the UV spectrum. It detects ultraviolet (UV) radiation at wavelengths of 280 to 360 nanometers. The spectral response is closely matched to the Erythema Action Spectrum, defined by McKinlay and Diffey (1987) and internationally recognized as the radiation that is most responsible for causing redness of the human skin.
Specifications
- Operating temperature: -40 °F to +150 °F (-40 °C to +65 °C)
- Storage temperature: -50 °F to +158 °F (-45 C° to +70 °C)
- Transducer: Semiconductor photodiode
- Spectral response: 280 to 360 nm (Erythema Action Spectrum)
- Cosine response: ±4% FS (0° to 90° zenith angle)
- Supplied cable length: 2′ (0.6 m)
- Cable type: 4-conductor, 26 AWG
- Connector: Modular RJ-11
- I/O Specs
- Green wire: Output (0 to 2.5Vdc); 150 mV per UV Index, 364 mV per MED/hour
- Black & red wires: Ground
- Yellow wire: +3V ±10%, 2.4 mA
- Housing material: UV-resistant ABS plastic
- Dimensions (length x width x height): 2″ x 2.75″ x 2.25″ (51 mm x 70 mm x 57 mm)
- Weight: 0.5 lbs (226 g)