230-61002 Gill Pressure Port
The 230-61002 Gill Pressure Port provides an effective static pressure inlet for a barometric pressure sensor. The unique parallel plate design effectively slows air velocity at the pressure inlet. A simple internal baffle system creates a barrier to water and snow intrusion into the barometer.
The 230-61002 Gill Pressure Port provides an effective static pressure inlet for a barometric pressure sensor. Measurement performance of remote barometers can be seriously impaired by the effects of wind at the barometer inlet. Ambient wind of 20 meters/second blowing over a typical barometer inlet tube can cause dynamic pressure errors as high as 3 millibars.* These errors can make atmospheric pressure data useless for research and forecast purposes.
The Gill Pressure Port reduces dynamic pressure errors to a minimum level. The unique parallel plate design of the Pressure Port effectively slows air velocity at the pressure inlet. A simple internal baffle system creates a barrier to water and snow intrusion into the barometer. The Gill Pressure Port features durable, corrosion-resistant plastic construction for long service and minimum cost.
Barbed fittings are supplied for easy connection using 1/4 inch i.d. flexible tubing. (3.3 feet of tubing is supplied.) The installation procedure is as follows:
- Select a location that is away from sources of contamination. For marine use, the pressure port should be located as high above the water as practical. For land installations, dirty and dusty locations should be avoided.
- Securely attach pressure port to vertical pipe using U-bolt mounting. Connect 1/4 inch flexible tubing to barbed fitting.
- Route flexible tubing to barometric pressure sensor location. Connect to the barbed fitting on the pressure sensor. The pressure port is designed to exclude moisture. In use, however, condensation in the flexible tubing may cause water to collect. If this occurs, the tubing should be drained to prevent blockage. A T-fitting and sump may be installed. If allowed to accumulate, water blockage in the tubing will cause incorrect pressure readings.
* Development and Testing of a No-Moving-Parts Static Pressure Inlet for Use on Buoys -Gerald C. Gill
Specifications
- Dynamic Pressure Error: 0.5 mb max. @ 20 m/s
- Dimensions: Height: 4.5″ (11.5 cm); Diameter: 5.1″ (13 cm)
- Mounting: Offset bracket with U-Bolt
- Weight: 0.4 lb (0.2 kg)
Downloads
- 230-61002 Data Sheet
- 230-61002 Manual – available on request
Shipping Info
Important Notes About Shipping Charges
Unless specific instructions accompany the order, we will use our judgment as to the best method of shipment. Most orders are shipped via United Parcel Service. Special arrangements can be made for other carriers. All shipments are F.O.B. Grass Valley, California. Title and risk of loss will pass to the purchaser upon delivery to the freight carrier. The purchaser is responsible for all transportation and insurance charges. Shipping containers are not returnable.
The shipping charges provided by our website shopping cart are estimated based on the individual items in your order. If you are in the United States and have ordered more than one item, the actual shipping charge will sometimes be less than the amount quoted. If we have overcharged you for shipping, we will refund the difference after your order is shipped. If the actual shipping charge is higher than the amount quoted, we will contact you before shipping your order.
If you have any questions or special requirements, or to request an exact freight quote, please send a message or call us at +1 (530) 823-7185.